These guidelines should be used as initial starting points only, and may be changed as needed to satisfy specific site conditions and sensitometric results. Replenishment Rates -- Dedicated Mammography Note: A processor is considered dedicated if only single-emulsion film mammography, ultrasound, etc. If using Replenisher Tank: Replenisher tank should be sized such that volume is approximately equal to or less than, the volume used in 2 weeks. Mix only in quantities large enough to be used in 2 weeks or less.

If using an Automixer without a floating lid: Mix only in quantities large enough to be used in 1 week. Mix only when replenisher volume is near or at the low level alarm. Note: Do not feed film into the processor during chemical mixing. For specific processors, refer to the General Processor Information.

Starter Volumes. Processor Maintenance Timing Drain, clean, and refill the developer and fixer processing tanks monthly or as experience indicates. Specific site conditions may dictate more or less cleaning. Be sure to follow all environmental regulations when disposing of processing solutions. This amount will be fed into the processor every 5 minutes.

Once set, the rate may be changed depending on the individual circumstances. Monitoring the processor sensitometry is required to change replenishment rates. Use the following recommendations as replenishment rates are reduced Monitor sensitometry speed and contrast.

Reduce developer and fixer replenishment rates by 5 ml. Monitor sensitometry for 2 weeks If no change is seen, rates may be reduced by another 5 ml.

Once a change is seen, increase the developer and fixer rates by 5 to 10 ml. Set the flooded replenishment rate at one-half of the above amount. The processor will feed the one-half amount in this case, 50 ml into the processor every 20 minutes. Monitoring the processor sensitometry is required to reduce replenishment rates. Use the following recommendations as replenishment rates are reduced: Monitor sensitometry speed and contrast. If you are not sure whether the Processor is set to Flooded or Regular Mode, ask your service provider.

November — 9B Page Preventive Maintenance Preventive Maintenance Section 4: Preventive Maintenance Important For trouble-free operation of the Processor, perform maintenance periodically in accordance with the following suggestions. Weekly Maintenance Caution When you install or remove the Fixer Rack 15 , be careful not to splash fixer chemicals into the Developer Tank Page Warranty U. Only Kodak.

This manual is also suitable for: X-omat j X-omat a. Print page 1 Print document 16 pages. Rename the bookmark. Delete bookmark? Cancel Delete. Delete from my manuals? Sign In OR. Don't have an account? Sign up!